Air Quality in the IVF Laboratory
Rada Medical

Air Quality in the IVF Laboratory: The Hidden Environmental Factor in Embryo Development

In vitro fertilization (IVF) depends on an exceptionally controlled biological environment. Temperature, pH, osmolality, gas concentrations, culture media, incubation conditions, and handling procedures are routinely monitored because even relatively small deviations can affect gametes and embryos.

One environmental factor, however, is sometimes less visible in day-to-day embryology practice: the air surrounding the laboratory, workstations, and incubators.

For an IVF laboratory, “clean air” means more than simply removing dust or preventing microbial contamination. Airborne particles, volatile organic compounds (VOCs), chemical pollutants, microorganisms, and contaminants introduced through construction materials, cleaning products, equipment, personnel, or the building's ventilation system can all become part of the laboratory environment.

Current recommendations from the European Society of Human Reproduction and Embryology (ESHRE), together with international consensus documents and guidance from the American Society for Reproductive Medicine (ASRM), emphasize that air quality should be actively controlled, monitored, and incorporated into the overall quality-management system of the IVF laboratory.

 

Why Does Air Quality Matter in an IVF Laboratory?

Human embryos are particularly sensitive to their physical and chemical environment during in vitro culture.

Inside the body, an embryo develops within a highly regulated biological system. In the laboratory, many of those environmental conditions must be recreated artificially. The embryo is therefore indirectly exposed to the laboratory environment through the culture system.

Air contaminants can reach embryos through several pathways. They may enter the incubator, interact with culture media or mineral oil, contaminate work surfaces, or be introduced during procedures in which gametes and embryos are exposed to the surrounding atmosphere.

Research and expert consensus have associated poor laboratory air quality—particularly chemical contamination—with impaired embryo development and, in some studies, poorer implantation and pregnancy outcomes. At the same time, the evidence is heterogeneous, and randomized clinical evidence isolating the effect of individual air contaminants is difficult to obtain.

This distinction is important: air quality should be treated as a controllable risk factor in IVF laboratory quality management, but it should not be presented as a single variable that determines IVF success.

 

The Two Major Air-Quality Challenges: Particles and Chemicals

A useful way to approach IVF laboratory air quality is to divide airborne contamination into two broad categories.

  1. Particulate contamination

Particles may include dust, skin and hair fragments, textile fibers, droplets, microorganisms, and other airborne material.

Particulate contamination is important because particles can carry microorganisms and other contaminants. It is also an indicator of how effectively the laboratory controls personnel movement, airflow, filtration, cleaning, and ingress of contaminants.

ISO 14644-1 provides an internationally recognized framework for classifying cleanroom air cleanliness according to airborne particle concentration. Importantly, ISO 14644-1 addresses particle concentration; it does not by itself characterize the chemical toxicity, viable contamination, or biological nature of those particles.

For IVF laboratories, therefore, particle classification is only one component of environmental control.

  1. Chemical contamination and VOCs

Volatile organic compounds are often the more difficult—and potentially more overlooked—problem.

VOCs are chemicals that can evaporate into the air under normal indoor conditions. Potential sources in an IVF facility include:

  • Paints and coatings
  • Adhesives and sealants
  • Flooring and furniture
  • Construction and renovation materials
  • Cleaning and disinfecting products
  • Air fresheners and fragrances
  • Cosmetics and personal-care products
  • Laboratory consumables
  • Packaging materials
  • Plastics and polymers
  • Equipment and electrical components
  • Building materials outside the laboratory
  • Vehicle exhaust or industrial pollution entering through HVAC fresh-air intakes

The 2018 Cairo Consensus on IVF laboratory air quality specifically identified construction materials, cleaning products, equipment, consumables, personnel-related contaminants, and HVAC systems as potential sources of chemical contamination.

This is why an IVF laboratory can have apparently clean air from a particulate perspective while still having a chemical air-quality problem.

 

VOCs: The Hidden Risk

VOCs are particularly relevant to embryo culture because the laboratory environment interacts with the culture system.

An IVF laboratory is effectively a controlled interface between air, equipment, culture media, and living cells. Chemicals present in ambient air can potentially enter incubators or contact materials used for embryo culture.

Some VOCs, particularly aldehydes and other reactive compounds, are of concern because of their potential biological toxicity.

The Cairo Consensus concluded that there is sufficient evidence from animal and human studies to justify substantial efforts to control laboratory contamination, while also recognizing that high-level clinical evidence is difficult to generate.

The practical implication is straightforward:

Do not wait for a measurable decline in embryo development before investigating air quality.

Environmental control should be preventive rather than reactive.

 

Where Do IVF Laboratory Air Contaminants Come From?

A comprehensive environmental risk assessment should consider the entire laboratory ecosystem rather than only the room containing the incubators.

Construction and renovation

New construction and renovation can introduce substantial amounts of VOCs.

Fresh paint, flooring, adhesives, sealants, cabinetry, furniture, ceiling materials, and other building products can continue releasing chemicals after installation.

The current ESHRE recommendations specifically advise that materials used in laboratory construction, painting, flooring, and furniture should be appropriate for cleanroom conditions and selected to minimize VOC release and potential embryo toxicity. ESHRE also recommends allowing sufficient time for construction materials to off-gas before a newly constructed or renovated laboratory is brought into clinical operation.

A laboratory renovation should therefore be treated as an environmental qualification event—not simply as a construction project.

Cleaning and disinfection

Infection control is essential, but the products used to achieve it can themselves become sources of chemical contamination.

Disinfectants, detergents, solvents, aerosols, fragrances, and other cleaning products can release airborne chemicals.

The objective should not be to minimize cleaning. Instead, IVF laboratories should establish validated cleaning procedures that balance microbiological control with chemical safety.

Cleaning products should be selected based on:

  • Toxicological profile
  • VOC emissions
  • Residue characteristics
  • Compatibility with laboratory surfaces
  • Required contact time
  • Required concentration
  • Ventilation requirements
  • Potential proximity to gametes and embryos

Cleaning activities involving strong chemicals should be controlled and, where possible, physically separated from critical embryo-handling areas.

 

Personnel

People are another source of contamination.

Skin particles, hair, fibers, cosmetics, fragrances, personal-care products, and clothing can contribute to airborne contamination.

This is one reason controlled access, appropriate laboratory clothing, hand hygiene, and behavioral procedures are important components of IVF laboratory environmental control.

The updated ESHRE recommendations include restricted laboratory access and appropriate changing and hand-washing arrangements among good laboratory practices.

 

Consumables and packaging

Consumables should not be assumed to be chemically inert simply because they are manufactured for laboratory use.

Packaging materials, plastics, tubing, adhesives, and other components may release VOCs, particularly when new.

Where appropriate, laboratories should evaluate new materials before introducing them into critical areas and should establish procedures for storage and, where necessary, off-gassing.

 

HVAC and outdoor pollution

The HVAC system can either protect the laboratory or introduce contaminants.

Outdoor air may contain traffic-related pollutants, industrial emissions, smoke, construction-related chemicals, or other contaminants. The location of the outdoor air intake therefore matters.

The Cairo Consensus recommends assessing the location of the laboratory and potential external pollution sources during planning and design.

This means that air-quality management begins before air enters the laboratory.

 

HEPA Filtration Is Necessary—but Not Sufficient

HEPA filtration is highly effective for controlling airborne particles. However, a HEPA filter should not be regarded as a complete solution for IVF laboratory air quality.

HEPA filtration primarily addresses particulate contamination. VOCs are gaseous chemical contaminants and require a different control strategy.

Current ESHRE recommendations specifically call for both HEPA and VOC control in IVF laboratory air systems.

Depending on the system design, VOC control can involve technologies such as activated carbon and other chemical filtration media. The selection should be based on the actual contaminants, airflow requirements, filter specifications, maintenance requirements, and laboratory risk assessment.

ASRM guidance similarly recommends dedicated clean air for embryology laboratories, with consideration of HEPA filtration as well as carbon and permanganate filtration to address particles, VOCs, and inorganic air pollutants.

The practical lesson is:

Particle filtration and chemical filtration solve different problems. A robust IVF air-management system should address both.

 

Positive Pressure: Keeping Contaminants Out

An IVF laboratory should generally be maintained at positive pressure relative to surrounding spaces.

Positive pressure means that air flows outward when doors or other openings are opened, helping prevent uncontrolled air from adjacent areas from entering the laboratory.

ESHRE recommends positive pressure to minimize air contamination, while ASRM guidance also identifies positive laboratory pressure relative to adjacent areas as an important design consideration.

Pressure relationships should be designed as part of the overall airflow strategy rather than treated as an isolated number.

A laboratory should establish:

  • Pressure relationships between critical and non-critical rooms
  • Alert and action levels
  • Continuous or periodic monitoring appropriate to risk
  • Alarm procedures
  • Response procedures for pressure failure
  • Maintenance and calibration requirements

 

Air Changes: More Is Not Always Better

A common misconception is that the safest IVF laboratory is simply the one with the highest possible air-exchange rate.

That is not necessarily correct.

Excessive airflow can create unwanted temperature fluctuations, increase energy demand, disturb the laboratory environment, and potentially interfere with the stability required for gametes and embryos.

The Cairo Consensus proposed 15 total air changes per hour, including approximately three fresh-air changes per hour, as a design benchmark for new ART laboratories. However, these figures should be interpreted as consensus-based aspirational benchmarks rather than a universal prescription for every laboratory.

The appropriate HVAC design should consider:

  • Laboratory size
  • Number of personnel
  • Workload
  • Heat generated by equipment
  • Incubator configuration
  • Air filtration
  • VOC-control technology
  • Fresh-air requirements
  • Pressure cascade
  • Temperature and humidity
  • Local regulations
  • Laboratory layout
  • Risk assessment

The objective is controlled and stable air—not simply maximum airflow.

 

Monitoring Air Quality: What Should an IVF Laboratory Measure?

Air quality management is not complete when an HVAC system is installed.

The laboratory should establish an environmental monitoring program with defined parameters, sampling locations, frequencies, alert levels, action levels, responsibilities, and corrective actions.

The current ESHRE recommendations state that IVF laboratory air cleanliness should be assessed using a risk-based approach at least annually, both in “at rest” and “in operation” conditions.

A practical monitoring program can include:

Particle monitoring

Particle counts provide information about airborne particulate cleanliness.

Monitoring should be designed around the laboratory's risk profile and applicable cleanroom classification requirements.

ISO 14644-2 emphasizes the importance of a monitoring plan and ongoing trend analysis rather than relying only on occasional classification testing.

VOC monitoring

VOC monitoring deserves particular attention in IVF laboratories.

Where possible, laboratories should consider measuring specific VOCs rather than relying exclusively on a single total-VOC measurement. The Cairo Consensus specifically highlighted the importance of investigating individual VOCs because different compounds have different biological and chemical properties.

Monitoring should be used to establish:

  • Baseline conditions
  • Normal operating range
  • Trends over time
  • Effects of renovation
  • Effects of new equipment or consumables
  • Effects of cleaning procedures
  • Unexpected environmental events

A sudden increase in VOCs should trigger an investigation rather than simply being recorded.

Microbiological monitoring

Airborne and surface microbiological monitoring should form part of the laboratory's contamination-control strategy.

However, microbiological cleanliness and chemical cleanliness are different parameters. A laboratory can pass microbial monitoring while still experiencing unacceptable chemical contamination.

The two systems therefore need to be managed independently.

Pressure monitoring

Pressure differentials should be monitored according to the laboratory's HVAC design and risk assessment.

An unexpected pressure loss can indicate:

  • HVAC malfunction
  • Door or seal failure
  • Filter loading
  • Fan problems
  • Building changes
  • Air-balance problems

Pressure monitoring should therefore be connected to an established investigation and corrective-action process.

 

Practical IVF Laboratory Air-Quality Checklist

For an embryology laboratory, the following checklist provides a useful starting point for an environmental risk assessment.

Laboratory design

  • Use low-emission or low-VOC construction materials.
  • Avoid unnecessary sources of adhesives, sealants, fragrances, and chemical emissions.
  • Design controlled access to the embryology laboratory.
  • Establish appropriate pressure cascades.
  • Keep high-risk contamination sources outside critical laboratory areas.
  • Position HVAC air intakes away from obvious external pollution sources.
  • Separate cleaning/sterilization activities from embryo-handling areas where possible.
  • Design the laboratory so that airflow can be validated and monitored.

HVAC system

  • Use dedicated or appropriately controlled air handling.
  • Include HEPA filtration where required by the laboratory design and applicable standards.
  • Include appropriate VOC filtration.
  • Maintain positive pressure relative to adjacent areas.
  • Control temperature and humidity.
  • Maintain the HVAC system continuously as required by the validated laboratory design.
  • Establish preventive maintenance and filter-replacement schedules.
  • Monitor pressure and airflow performance.
  • Validate the system after installation, modification, or major maintenance.

VOC control

  • Establish a baseline VOC profile.
  • Investigate sources rather than relying only on filtration.
  • Control VOC-emitting construction materials.
  • Evaluate new consumables and equipment.
  • Control cleaning and disinfectant products.
  • Minimize unnecessary fragrances and aerosols.
  • Allow adequate off-gassing after construction or renovation.
  • Perform environmental testing before returning a renovated laboratory to clinical use.
  • Trend results over time.

Personnel and operational practices

  • Restrict access to authorized personnel.
  • Use appropriate laboratory clothing.
  • Minimize unnecessary movement and door opening.
  • Control cosmetics and fragrances.
  • Store chemicals and cleaning products appropriately.
  • Keep non-essential materials out of critical areas.
  • Establish documented procedures for responding to air-quality deviations.

Monitoring and quality management

  • Establish written environmental specifications.
  • Define alert and action levels.
  • Monitor particles according to the laboratory's risk-based plan.
  • Monitor VOCs using an appropriate validated method.
  • Monitor pressure relationships.
  • Conduct microbiological environmental monitoring.
  • Review trends rather than individual results alone.
  • Investigate excursions using root-cause analysis.
  • Document corrective and preventive actions.
  • Requalify the laboratory following significant construction, HVAC modification, or other environmental changes.

 

What Should Happen After IVF Laboratory Renovation?

Renovation is one of the highest-risk periods for chemical contamination.

A newly renovated laboratory should not automatically return to clinical use simply because construction has finished and surfaces have been cleaned.

A practical commissioning sequence is:

  1. Complete construction.
  2. Remove construction materials and waste.
  3. Perform intensive cleaning using validated procedures.
  4. Operate the HVAC and filtration system as appropriate to the commissioning plan.
  5. Allow sufficient time for off-gassing.
  6. Perform particle, VOC, pressure, temperature, humidity, and microbiological assessments as appropriate.
  7. Compare results with predefined acceptance criteria.
  8. Investigate any abnormal VOC or particle results.
  9. Verify HVAC performance and filtration.
  10. Approve the laboratory for clinical use only after environmental qualification is satisfactory.

The Cairo Consensus recommends allowing adequate time for off-gassing following construction or renovation, with two to three weeks identified as a minimum planning period in its consensus recommendations, while emphasizing that the actual time required depends on the materials and environment.

The updated ESHRE recommendations likewise emphasize adequate off-gassing time after construction or renovation.

 

What About VOC Target Values?

One frequently cited benchmark comes from the Cairo Consensus, which proposed:

  • Total VOCs: less than 500 µg/m³
  • Aldehydes: less than 5 µg/m³

These values are useful as reference points when designing or evaluating ART laboratory environments, but they should not automatically be presented as universal regulatory limits for every IVF laboratory. The consensus describes them as aspirational benchmarks, and local regulatory requirements may differ.

This distinction matters because laboratory air-quality specifications should be based on applicable regulations, accreditation requirements, engineering validation, risk assessment, and the clinical context.

 

Air Quality Should Be Part of the IVF Quality Management System

Air quality should not sit entirely within the responsibility of the facility or engineering department.

For an IVF laboratory, environmental conditions are directly connected to the biological processes performed there. Embryologists, laboratory managers, quality managers, engineers, infection-control personnel, and facility teams should therefore work together.

A mature environmental quality-management system should connect:

Environmental monitoring → deviation detection → investigation → root-cause analysis → corrective action → verification → trend analysis.

For example, if VOC levels suddenly increase, the response should not simply be “replace the filter.”

The laboratory should ask:

  • Was there recent construction?
  • Was a new cleaning product introduced?
  • Were new consumables purchased?
  • Was new furniture installed?
  • Was laboratory equipment serviced?
  • Did the HVAC system change operating conditions?
  • Was there external construction or pollution?
  • Was the sampling method appropriate?
  • Did the increase occur in one location or throughout the laboratory?
  • Did the change correlate with any laboratory process?

This approach converts air quality from a facility-maintenance issue into a measurable component of laboratory quality management.

 

Air Quality and Embryo Development: What Does the Evidence Actually Show?

The scientific literature provides a strong rationale for controlling laboratory contamination, but it is important not to overstate the evidence.

Experimental studies have demonstrated that certain airborne contaminants can adversely affect embryo development. Human observational studies have also reported associations between improved laboratory air quality and IVF laboratory outcomes.

However, the clinical literature has limitations.

It is difficult to conduct randomized trials in which embryos are deliberately exposed to different laboratory atmospheres. Air quality also interacts with many other variables, including culture media, incubators, laboratory technique, operator performance, temperature, gas composition, and embryo-selection practices.

Consequently, air quality should be understood as one component of a highly interconnected embryo-culture system.

The absence of perfect clinical evidence is not a reason to ignore environmental control. Rather, it supports the use of risk-based quality management and continuous environmental monitoring.

 

The Modern Approach: From “Clean” to “Controlled”

The concept of a clean IVF laboratory has evolved.

Historically, attention was often concentrated on visible cleanliness and airborne particle counts. Modern embryology requires a broader approach that considers:

Particles + microorganisms + VOCs + airflow + pressure + temperature + humidity + materials + cleaning chemicals + consumables + personnel + monitoring.

ESHRE's current recommendations explicitly incorporate HEPA filtration, VOC control, positive pressure, controlled environments, regular monitoring, and risk-based assessment of laboratory air cleanliness.

At the same time, ISO cleanroom standards provide useful frameworks for particle classification and monitoring, while recognizing that particle cleanliness does not automatically characterize chemical or microbiological contamination.

This distinction is particularly important in IVF because the laboratory is not simply protecting a manufactured product. It is supporting the culture of living human cells during a critical stage of early development.

 

Final Takeaway

Air quality is one of the least visible components of IVF laboratory performance, but it should not be treated as a secondary consideration.

A well-designed embryology laboratory should control both particulate and chemical contamination. HEPA filtration can address airborne particles, while appropriate VOC-control strategies are necessary for chemical contaminants. Positive pressure, appropriate HVAC design, low-emission construction materials, controlled cleaning practices, restricted access, and systematic environmental monitoring complete the framework.

Most importantly, air quality should be managed proactively.

The objective is not merely to have a laboratory that looks clean. The objective is to create a stable, validated, monitored environment in which gametes and embryos are protected from avoidable physical, chemical, and biological stress.

For IVF laboratories seeking to improve environmental control, the practical starting point is straightforward: establish a baseline, identify contamination sources, validate the HVAC and filtration system, define measurable acceptance criteria, monitor trends, and treat every significant environmental change as a potential quality event.

In embryo culture, what cannot be seen can still matter.

 

 

References

  1. Arroyo G, Barrie A, Coticchio G, et al. ESHRE recommendations on Good Practice in the IVF laboratory. Human Reproduction. 2026;41(8):1245–1287. DOI: 10.1093/humrep/deag096.
  2. ESHRE Good Practice in the IVF Laboratory Working Group. ESHRE Recommendations on Good Practice in the IVF Laboratory. European Society of Human Reproduction and Embryology, 2026.
  3. de los Santos MJ, Apter S, Coticchio G, et al. Revised guidelines for good practice in IVF laboratories (2015). Human Reproduction. 2016;31(4):685–686. DOI: 10.1093/humrep/dew016.
  4. Mortimer D, Cohen J, Mortimer ST, et al. Cairo consensus on the IVF laboratory environment and air quality: report of an expert meeting. Reproductive BioMedicine Online. 2018;36(6):658–674. DOI: 10.1016/j.rbmo.2018.02.005.
  5. Morbeck DE. Air quality in the assisted reproduction laboratory: a mini-review. Journal of Assisted Reproduction and Genetics. 2015;32:1019–1024.
  6. Esteves SC, Bento FC. Air quality control in the ART laboratory is a major determinant of IVF success. Journal of Assisted Reproduction and Genetics. 2016.
  7. Agarwal A, et al. Volatile organic compounds and good laboratory practices in the in vitro fertilization laboratory: the important parameters for successful outcome in extended culture. Journal of Assisted Reproduction and Genetics. 2017.
  8. American Society for Reproductive Medicine. Comprehensive guidance for human embryology, andrology, and endocrinology laboratories: management and operations: a committee opinion. Fertility and Sterility. 2022;117:1183–1202.
  9. American Society for Reproductive Medicine. Minimum standards for practices offering assisted reproductive technologies: a committee opinion. 2021.
  10. ISO. ISO 14644-1:2015. Cleanrooms and associated controlled environments — Part 1: Classification of air cleanliness by particle concentration. International Organization for Standardization.
  11. ISO. ISO 14644-2:2015. Cleanrooms and associated controlled environments — Part 2: Monitoring to provide evidence of cleanroom performance related to air cleanliness by particle concentration. International Organization for Standardization.
  12. Cairo Consensus Group. ‘There is only one thing that is truly important in an IVF laboratory: everything’ Cairo Consensus Guidelines on IVF Culture Conditions. Reproductive BioMedicine Online. 2020;40(1):33–60.
Sharing in:
Rada Medical

RADA Medical

RADA, a brand owned by Behrad Rooyesh Royan Company, is a medical device manufacturer dedicated to supporting individuals and families on their journey to parenthood and advancing women’s health. Through safe and effective solutions for assisted reproduction and gynecological care, RADA empowers healthcare professionals to deliver reliable and high-quality patient care.